| Model Number |
7K78-35 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer observed a falsely elevated b-hcg result on the architect i2000sr analyzer.The following data was provided: initial 724.61 miu/ml, repeat < 1.2 miu/ml.There was no impact to patient management reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of field data, and a review of labeling.No adverse trend was identified for the customer's issue.Historical performance of the reagent lot was evaluated using world wide data.The patient data was analyzed and within established control limits.Labeling was reviewed and found to be adequate.Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
|
| |
|