• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. FMP HIP; LINER, ACETABULAR, FMP CONSTRAINED 28XMP9

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. FMP HIP; LINER, ACETABULAR, FMP CONSTRAINED 28XMP9 Back to Search Results
Model Number 436-28-009
Health Effect - Clinical Code No Code Available (3191)
Date of Event 11/20/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Second revision surgery - the surgeon performed a poly exchange due to the patient having soft tissue loosening.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to soft tissue loosening.The previous surgery and the revision detailed in this investigation occurred 2.7 months apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to the soft tissue loosening.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the patient's instability.There are multiple factors that may contribute to the event that are outside the control of djo surgical such as: patient activities, loose joints from degenerative tissues, excessive loads and range of motions, patient non-compliance with medical instructions or trauma.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMP HIP
Common Device Name
LINER, ACETABULAR, FMP CONSTRAINED 28XMP9
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7096685
Report Number1644408-2017-01147
Device Sequence Number11246169
Product Code KWZ
UDI-Device Identifier00888912085731
UDI-Public(01)00888912085731
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K023794
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/17/2021
Device Model Number436-28-009
Device Catalogue Number436-28-009
Device Lot Number501C1012
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/20/2017
Supplement Date Received by Manufacturer01/19/2018
Initial Report FDA Received Date12/07/2017
Supplement Report FDA Received Date01/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
436-28-000, LOT 496C1064
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
-
-