| Model Number |
FG-000150 |
| Medical Device Problem Codes |
Retraction Problem (1536); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
11/07/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was retained by the user facility and was not returned to the manufacturer for evaluation.Poor quality photographs of the device were obtained which did not allow an adequate investigation of the device.In discussion with the surgeon performing the procedure, the procedure was completed as intended and the device functioned properly.The surgeon believes the metallic component (spring) was caught on the trocar during removal of the device and does not recommend another procedure to recover the component.The surgeon does not believe this was a device performance issue but rather use error.
|
| |
|
Event or Problem Description
|
|
The patient underwent a robotic tecab procedure in which the c-port device was used to perform the distal bypass anastomosis.The device functioned as intended and the procedure was completed as intended.Following a post-operative chest x-ray, a small metallic density was noted in the patient's left chest that was previously not recognized.After surgery, the patient was transferred for normal postoperative care.No patient complications have been reported as a result of the metallic density.
|
| |
|