| Model Number |
00711784 |
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product was not returned to us endoscopy.The lot history record was reviewed and found no issues during manufacture; the device was built to specification.The instructions for use contain the following warnings and precautions: "exposure to bodily fluids may occur during connection or disconnection of these devices; adherence to body substance isolation protocols is the responsibility of the user.Do not leave a device hanging from the valve.Doing so can cause the creation of a larger valve slit/hole that may cause leakage.If the lid of the valve is opened while attached to the endoscope during a procedure, scope suction will be compromised and leakage may occur.If leakage occurs, a sterile gauze should be used to cover the valve." in-service training has been provided to the user.
|
| |
|
Event or Problem Description
|
|
The bioshield biopsy valve is used to cover the opening to the biopsy/suction channel inlet of a gastrointestinal endoscope.The bioshield biopsy valve provides access for endoscopic device passage and exchange, helps maintain insufflation and minimizes leakage of biomaterial from the biopsy port throughout the gastrointestinal endoscopic procedure.The user facility reported that post procedure, the biopsy valve in the bioguard air/water and suction valve kit sprayed fluid.There was no contact with patient or user.
|
| |
|
Search Alerts/Recalls
|