MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD BIOSYN; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD BIOSYN; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID Back to Search Results
Model Number UM-877
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/15/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
To date, the incident sample has not been received for evaluation.If the sample is received, or if additional information pertinent to the incident is obtained, a follow-up report will be submitted.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the needle detached from the thread during a procedure.Nothing fell into the cavity.Another device was used to complete the procedure.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned products noted suture and needle.The needle was observed to be detached from the suture.Upon microscopic inspection of the needle, normal needle crimp and swage marks were observed on the needle.The retainer was inspected with no abnormalities.As no sealed products were returned, functional testing was precluded.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Should new information become available, the file will be re-opened and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the needle detached from thread.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOSYN
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key7099746
Report Number9612501-2017-06740
Device Sequence Number1235297
Product Code GAM
UDI-Device Identifier20884521033457
UDI-Public20884521033457
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K000037
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2022
Device Model NumberUM-877
Device Catalogue NumberUM-877
Device Lot NumberD7E1794X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2017
Initial Date Received by Manufacturer 11/17/2017
Supplement Date Received by Manufacturer02/07/2018
Initial Report FDA Received Date12/08/2017
Supplement Report FDA Received Date02/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-