BARD PERIPHERAL VASCULAR, INC. TRUE BALLOON AORTIC VALVULOPLASTY DILATATION CATHETER Back to Search Results
Catalog Number 0204511
Medical Device Problem Codes Deflation Problem (1149); Aspiration Issue (2883)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The return of the device is pending.The results of the anticipated device evaluation will be provided upon completion of the event investigation.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during preparation for a tavi procedure, the valvuloplasty balloon allegedly could not be deflated with a syringe.Patient status is unknown.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Conclusion: the device was returned for evaluation.A visual inspection did not find any aspiration related defects.The device was prepped per the instructions for use, and no air was seen aspirating into the device.Therefore, the investigation is unconfirmed for the reported aspiration issue.The definitive root cause for the reported aspiration issue could not be determined based upon the available information.It is unknown if procedural issues contributed to the reported event.Labeling review: the current instructions for use (ifu) states: warnings & precautions: never use force when advancing or retracting intravascular device without first angiographically determining cause for any resistance.Using force may damage catheter or cause injury to the patient (such as vessel perforation).Do not exceed maximum inflation pressure indicated on label.Excess inflation pressure can cause balloon rupture and the inability to withdraw catheter through introducer sheath.Withdrawing balloon through introducer sheath may damage balloon.Balloon deflation and catheter removal: deflate balloon by drawing vacuum on the = 50 cc inflation device.Use fluoroscopy to visually confirm that balloon has fully deflated prior to withdrawing catheter.While maintaining negative pressure and guidewire position, withdraw deflated device over the guidewire and out through introducer sheath.Use of a gentle clockwise twisting motion may be used to help withdraw balloon through introducer sheath.If unusual resistance is met when attempting to withdraw balloon, position balloon in anatomical position in which it can be inflated safely.Partially inflate balloon and then deflate it.Balloon re-folding can be observed under fluoroscopy.Use of recommended contrast concentration will facilitate fluoroscopic visualization of balloon re-folding.If balloon will not pass through introducer sheath for removal, remove balloon and sheath as a single unit.
 
Event or Problem Description
It was reported that during preparation for a tavi procedure, the valvuloplasty balloon allegedly could not be aspirated with a syringe.Patient status is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUE BALLOON AORTIC VALVULOPLASTY DILATATION CATHETER
Common Device Name
BALLOON AORTIC VALVULOPLASTY
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7102433
Report Number2020394-2017-01720
Device Sequence Number19057570
Product Code OZT
UDI-Device Identifier00801741090998
UDI-Public(01)00801741090998
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K150667
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/28/2020
Device Catalogue Number0204511
Device Lot NumberGFBS2599
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2017
Initial Date Received by Manufacturer 11/13/2017
Supplement Date Received by Manufacturer02/28/2018
Initial Report FDA Received Date12/08/2017
Supplement Report FDA Received Date03/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-