• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US ZT TRIAL SLEEVE 22D LRG; TRIAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US ZT TRIAL SLEEVE 22D LRG; TRIAL INSTRUMENT Back to Search Results
Catalog Number 257622115
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/11/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that all of the 10 srom instruments were deemed to be unusable because the size markings are no longer visible.
 
Additional Manufacturer Narrative
(b)(4).Investigation summary examination of the returned instrument confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary if information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZT TRIAL SLEEVE 22D LRG
Common Device Name
TRIAL INSTRUMENT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key7103841
Report Number1818910-2017-51131
Device Sequence Number11045596
Product Code LXH
UDI-Device Identifier10603295146919
UDI-Public10603295146919
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 11/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number257622115
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2017
Initial Date Received by Manufacturer 11/11/2017
Supplement Date Received by Manufacturer06/26/2018
Initial Report FDA Received Date12/11/2017
Supplement Report FDA Received Date07/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-