| Catalog Number |
Y917H |
| Medical Device Problem Code |
Detachment Of Device Component (1104)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested and the following was obtained: did the needle break into 2 pieces at the swage location? or did the needle pull away / separate from the suture ? did the needle fall into patient? if yes, was it removed? and how? any patient consequences / ae report? patient current status? (b)(6)rep confirmed via phone that the needle came away from the suture; the needle did not break.No broken tip was left in situ.No patient consequence.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Please clarify the date of the procedure/ event date.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent total knee replacement on unknown date and suture was used.The needle came away from swage after taking a bite of intact tissue.Another suture was used to complete the procedure.No patient consequence reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.A winding former, one paper lid and a detached needle were returned for analysis.However, the suture was not sent.During the visual inspection of the needle, the swage and attachment area of the needle were as expected.The barrel hole of the needle was examined under 20x magnification and no suture remnant was noted.No needle breakage was observed.The suture was not returned to determine the assignable cause.Per the condition of the needle, no defects or broken needle could be observed.Also, it could not be determined what may have caused the reported incident of ¿performance pull off suture needle¿.
|
| |
|
Search Alerts/Recalls
|