• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH, INC. TRUEMETRIX; BLOOD GLUCOSE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH, INC. TRUEMETRIX; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number TRUEMETRIX
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/16/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-20-user's test strip had poor storage (bathroom) (b)(4).
 
Event or Problem Description
Consumer reported complaint for high blood glucose test results.The customer is concerned with test results from results obtained of 199 mg/dl.The customer's expected fasting blood glucose test result range is 88 - 125 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2017, a back to back blood test was not performed by the customer.The product is not stored according to specification and is stored in the bathroom.The test strip lot manufacturer's expiration date is 01/31/2019 and open vial date is (b)(6) 2017.The meter memory was reviewed for previous test result history: result 1 :199mg/dl date:(b)(6) 2017 time:05:10 fasting, result 2 :77mg/dl date:(b)(6) 2017 time:15:20 fasting, result 3 :122mg/dl date:(b)(6) 2017 time:06:02 fasting, result 4 :141mg/dl date:(b)(6) 2017 time:11:54 fasting.
 
Additional Manufacturer Narrative
(b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-20-user's test strip had poor storage (bathroom).Test strip udi# (b)(4).
 
Event or Problem Description
Consumer reported complaint for high blood glucose test results.The customer is concerned with test results from results obtained of 199 mg/dl.The customer's expected fasting blood glucose test result range is 88 - 125 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2017, a back to back blood test was not performed by the customer.The product is not stored according to specification and is stored in the bathroom.The test strip lot manufacturer's expiration date is 01/31/2019 and open vial date is (b)(6) 2017.The meter memory was reviewed for previous test result history: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEMETRIX
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7107748
Report Number1000113657-2017-02051
Device Sequence Number19060583
Product Code NBW
UDI-Device Identifier00052569138526
UDI-Public(01)00052569138526
Combination Product (Y/N)N
Initial Reporter StateMO
PMA/510(K) Number
K140100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 12/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/31/2019
Device Model NumberTRUEMETRIX
Device Catalogue NumberRE4H01-81
Device Lot NumberMU2579
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2017
Was the Report Sent to FDA? No
Event Location Home
Initial Date Received by Manufacturer 11/16/2017
Supplement Date Received by Manufacturer11/16/2017
Initial Report FDA Received Date12/12/2017
Supplement Report FDA Received Date12/22/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-