• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SVCE REPL MDU CNTRL PWRMX ELITE; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SVCE REPL MDU CNTRL PWRMX ELITE; SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616SR
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/19/2017
Event Type  malfunction  
Manufacturer Narrative
The device was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Product failed functional testing with intermittent sticking oscillate button.Cause of sticky button is a corroded oscillate button spring.The root cause of sticky button has been determined to be corrosion of the oscillate button assembly.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Event Description
It was reported that the device was not engaging.No patient njury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVCE REPL MDU CNTRL PWRMX ELITE
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key7107951
MDR Text Key94671306
Report Number1643264-2017-01818
Device Sequence Number1
Product Code HAB
UDI-Device Identifier00885556582305
UDI-Public(01)00885556582305
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 12/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616SR
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/07/2017
Initial Date FDA Received12/12/2017
Supplement Dates Manufacturer Received12/12/2017
Supplement Dates FDA Received12/14/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-