Model Number 9500-44
Device Problem Component Missing (2306)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/13/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2017, that on (b)(6) 2017, a missing sensor wire occurred.The sensor was inserted into the abdomen on (b)(6) 2017.No medical intervention was reported.Additional event or patient information is not available.No product was provided for evaluation.The complaint confirmation was unable to be determined.A root cause could not be determined.
 
Manufacturer Narrative
(b)(4).
 
Event Description
A sensor was returned for evaluation.A visual inspection was performed and found that the wire is missing from the sensor pod and seal carrier.The customer's complaint of a missing sensor wire was confirmed.A root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key7108878
MDR Text Key94550061
Report Number3004753838-2017-111882
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Type of Report Initial,Followup
Report Date 11/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/14/2018
Device Model Number9500-44
Device Catalogue NumberSTS-GF-001
Device Lot Number5226692
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/14/2017
Initial Date FDA Received12/12/2017
Supplement Dates Manufacturer Received02/15/2018
Supplement Dates FDA Received03/16/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/14/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age63 YR
Patient Weight90
-
-