Sorin group (b)(4) manufactures the cobe revolution centrifugal blood pump with pc.The incident occurred in (b)(4).Per exemption number e2016005, (b)(4).The involved device was returned to sorin group (b)(4) for investigation.Visual inspection was unable to identify any defects.Simulated use testing was conducted and the device was found to emit an abnormal noise as the pump rotation was increased.However, this did not affect the performance of the device and the unit was able to deliver flow according to specification.As the flow issue could not be reproduced, a root cause was not identified.A possible cause of the noise could be internal mechanical damage to the hub or the impeller caused by rough handling, leading to anomalous rotation.
|