MEDTECH SA ROSA SURGICAL DEVICE COMPUTER-ASSISTED SURGICAL DEVICE Back to Search Results
Model Number ROSA 2.5.8
Device Problem Imprecision (1307)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/10/2017
Event Type  malfunction  
Manufacturer Narrative
According to results of the investigation, the root cause identified is a user error: the user did not report correctly the information provided by the rosa device on the electrode regarding the insertion distance. Investigation determined that the core device (b)(4) worked as intended. Not needed to identify root cause.
 
Event Description
It was reported that during per-operative imaging control the surgeon noticed that the electrode implanted was more than 2 cm far away from the target point planned. The root cause had been detected by the surgical team from a sliding of the stop of the electrode.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameROSA SURGICAL DEVICE
Type of DeviceCOMPUTER-ASSISTED SURGICAL DEVICE
Manufacturer (Section D)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR 34000
Manufacturer (Section G)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR 34000
Manufacturer Contact
jay sharma
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key7111870
MDR Text Key95737207
Report Number3009185973-2017-00823
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
PK101791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial
Report Date 12/13/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/13/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberROSA 2.5.8
Device Catalogue NumberN/A
Device Lot NumberN/A
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/13/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/07/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A

-
-