| Model Number |
SYM15F |
| Medical Device Problem Codes |
Degraded (1153); Product Quality Problem (1506)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/10/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic parastomal hernia repair procedure, while the mesh was being attached to the hernia defect, the device¿s mesh was damaged, wherein collagen film peeled off in the middle of the mesh in one area and the collagen was flaking around the outside.Saline was used to hydrate the mesh.Another piece of mesh was opened and used in order to resolve the issue.A 10/12 trocar was used.No patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The visual examination of the returned sample shows that the sample was returned in its original aluminum pouch.The sample was found contaminated by blood.The mesh diameter is around 10 cm and the blue marker is not placed in the middle of the mesh.The mesh was found torn from the left of the blue marker, until the bottom of the mesh.A blue yarn suture is found from the tear.A hole about 3 cm diameter was found in the left side of the marker.In one quarter of the mesh (under the blue marker until the tear), the collagen was found flaking on the edge and lacking in nearly all of this part.The investigation has concluded that excessive tension is highly suspected.If information is provided in the future, a supplemental report will be issued.
|
| |
|