• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THERAKOS XTS PHOTOPHERESIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THERAKOS XTS PHOTOPHERESIS SYSTEM Back to Search Results
Model Number NOT APPLICABLE
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/09/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction centrifuge bowl leak/break.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot f704 was conducted.There were no non-conformances associated with this lot.This lot met all release requirements.A review of f704 for the reported issue shows no trends.Trends were reviewed for complaint categories, noise, occlusion system alarm, leak centrifuge alarm, and centrifuge bowl leak/break.No trends were detected for each complaint category.This assessment is based on the information available at the time of the investigation.At the time of this report, the analysis of the customer provided photos is still in progress.A supplemental report will be filed when the analysis is complete.(b)(6).
 
Event or Problem Description
The customer called to report a centrifuge bowl leak/ break.During cycle 6 of the treatment, the customer heard a loud noise, following the loud noise sounded an occlusion system alarm.The alarm was silenced.From there a leak centrifuge alarm had occurred, this is when the blood was observed in the centrifuge and the surrounding areas.The treatment was aborted with no blood returned back to the patient.The patient was in stable condition and not affected by the incident.The customer has provided photographs to be investigated.
 
Additional Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction centrifuge bowl leak/break.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.Kit lot f704 was reviewed.There were no non-conformances related to the complaint.This lot met all release requirements.A review of kit lot f704 identified an increase in complaints associated with the centrifuge bowl leak/break complaint category.An investigation has been initiated in response to the increase of centrifuge bowl leak/break complaints observed with kit lot f704.Trends were reviewed for complaint categories, noise, occlusion system alarm, leak centrifuge alarm, and centrifuge bowl leak/break.No trends were detected for each complaint category.A photograph and video analysis was conducted for this complaint.A review of the customer provided photographs and video confirms the reported blood leak in the centrifuge chamber.However, the origin of the leak could not be determined based on the provided photographs, video and available information.A material trace used to build kit lot f704 found no related nonconformances.The device history record review did not result in any related non-conformances and this kit lot had passed lot release testing.The root cause of the following centrifuge bowl leak/ break could not be determined.No further action required.Investigation complete.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XTS PHOTOPHERESIS SYSTEM
Common Device Name
XTS PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
THERAKOS
bedminster NJ
MDR Report Key7112875
Report Number2523595-2017-00228
Device Sequence Number11271787
Product Code LNR
UDI-Device Identifier20705030200027
UDI-Public(01)20705030200027(10)F704(17)220301
Combination Product (Y/N)N
Initial Reporter CountryAR
PMA/510(K) Number
P860003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 01/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Nurse
Device Expiration Date03/01/2022
Device Model NumberNOT APPLICABLE
Device Catalogue NumberXT125
Device Lot NumberF704
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/14/2017
Supplement Date Received by Manufacturer01/10/2018
Initial Report FDA Received Date12/13/2017
Supplement Report FDA Received Date01/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age45 YR
Patient Weight63
-
-