MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 PREMIUM SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 PREMIUM SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9735665
Device Problems Computer Software Problem (1112); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/16/2017
Event Type  malfunction  
Manufacturer Narrative
No parts have been received by the manufacturer for evaluation.Part not returned for analysis.
 
Event Description
A site representative reported that, while in a spinal fusion, the cad model for the instrument would not appear on the application software when viewing saved plans nor the trajectory view.It was reported that the instrument appeared as a green line.Trajectory 2 and other views were reported to have shown the cad model without issue.The reported issue occurred at the sixth screw placement of the procedure and the surgeon completed the procedure without issue.There was a reported delay to the procedure of less than 1 hour due to this issue.The surgeon opted to complete the procedure without the use of the navigation system.There was no impact on patient outcome.No additional information was provided.
 
Manufacturer Narrative
The software investigation found that the reported event was related to a software issue.This issue was documented in a medtronic navigation software anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S8 PREMIUM SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
prashanth gali
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7114280
MDR Text Key94845815
Report Number1723170-2017-05116
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00643169722187
UDI-Public00643169722187
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/13/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9735665
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/16/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/04/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age72 YR
Patient Weight109
-
-