Model Number 173019
Medical Device Problem Code Detachment Of Device Component (1104)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Date of Event 11/17/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during surgery, the tip of the device came off and fell into the patient's cavity.It was immediately retrieved.Procedure was completed with a different device.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Evaluation summary post market vigilance (pmv) led an evaluation of one endo peanut 5mm device.Pmv observed that the cotton swab had disengaged from one of the shafts.No sign of adhesive or sheared cotton swab material was observed on the distal end of that shaft where the swab was affixed.A review of the device history record indicates this product was released meeting all quality release specifications at the time of manufacture.However, an assembly error was identified during product analysis.The most probable root cause of the condition maybe due to an inadequate amount of adhesive in affixing cotton tip.A process improvement has been implemented to prevent this condition from recurring.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO PEANUT
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key7114620
Report Number1219930-2017-09923
Device Sequence Number1379014
Product Code GCJ
UDI-Device Identifier20884523000761
UDI-Public20884523000761
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K931962
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Remedial Action Modification/Adjustment
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2022
Device Model Number173019
Device Catalogue Number173019
Device Lot NumberM7G02X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2017
Initial Date Received by Manufacturer 11/17/2017
Supplement Date Received by Manufacturer12/20/2017
02/14/2018
Initial Report FDA Received Date12/13/2017
Supplement Report FDA Received Date01/17/2018
03/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-