| Model Number |
173019 |
| Medical Device Problem Code |
Detachment Of Device Component (1104)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/17/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during surgery, the tip of the device came off and fell into the patient's cavity.It was immediately retrieved.Procedure was completed with a different device.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary post market vigilance (pmv) led an evaluation of one endo peanut 5mm device.Pmv observed that the cotton swab had disengaged from one of the shafts.No sign of adhesive or sheared cotton swab material was observed on the distal end of that shaft where the swab was affixed.A review of the device history record indicates this product was released meeting all quality release specifications at the time of manufacture.However, an assembly error was identified during product analysis.The most probable root cause of the condition maybe due to an inadequate amount of adhesive in affixing cotton tip.A process improvement has been implemented to prevent this condition from recurring.If information is provided in the future, a supplemental report will be issued.
|
| |
|