Model Number N/A
Device Problem Unsealed Device Packaging (1444)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/15/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Complaint sample was evaluated and the reported event was not confirmed.The product evaluation stated, "upon review of the packaging, the 01-97-6015a tyvek pouch, which is responsible for keeping the product sterile, showed evidence of a fully intact seal due to seal transfer marks across the pouch.Due to the fact that the tyvek pouch showed evidence of a full and complete seal, it is likely that the tyvek pouch was opened up in a severe distribution sequence, or some time after leaving zimmer biomet." dhr was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required as no were trends identified.Investigation results concluded that the reported event was due to transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.This report is being submitted late as it has been identified in remediation.
 
Event Description
When the nurse opened the packaging of the locking screw, it was noticed that the inner packaging wasn't sealed.Another screw was available and used without delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP LK SCR 3.5HEX 4.75X25 ST
Type of Device
PROSTHESIS, SHOULDER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7116999
MDR Text Key95596254
Report Number0001825034-2017-11184
Device Sequence Number1
Product Code KWS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK080642
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number180552
Device Lot Number288260
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/15/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/20/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-