• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME MONITOR Back to Search Results
Model Number XS PST
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/30/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
The customer stated that the coaguchek xs meter (b)(4) did not show complete segments in the three large "8s" seen in the results display.All of the top segments were missing, as well as some missing on the bottom and right sides.There was no physical damage evident with the display screen or meter.The meter was no dropped prior to the display issue.There was no incorrectly displayed result received or reported.The customer's meter was requested for investigation.The meter was returned for investigation.Upon initial investigation, a display check was performed on the meter and missing segments in the result field could be confirmed.No adverse events were alleged.
 
Manufacturer Narrative
Further investigations of the meter determined that the segment failures on the display occurred due to a defective contact of the conductive rubber.This was caused by twisted installation of the display within the frame.After correcting the position of the display and conductive rubber, the display showed no error.The display issue was a non-systematic error.No medical risk is likely.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key7117098
MDR Text Key95771627
Report Number1823260-2017-02927
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 01/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/14/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberXS PST
Device Catalogue Number03666794001
Is the Reporter a Health Professional? No
Date Manufacturer Received11/30/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-