Model Number TCN-10
Medical Device Problem Codes Break (1069); Material Discolored (1170); Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/11/2017
Type of Reportable Event Malfunction
Event or Problem Description
The device analysis performed on the returned electrode revealed discolored epoxy and a chip out.The color of newly cured epoxy is amber, light brown.If the epoxy is subjected to too many autoclave cycles it becomes brittle and discolored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NITINOL TC ELECTRODE
Common Device Name
ELECTRODES
Manufacturer (Section D)
COSMAN MEDICAL
22 terry ave
burlington MA 01803
MDR Report Key7118054
Report Number3006630150-2017-05266
Device Sequence Number16517916
Product Code GXI
UDI-Device Identifier00813250013054
UDI-Public00813250013054
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K050084
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 12/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Model NumberTCN-10
Device Lot Number110416
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/11/2017
Initial Report FDA Received Date12/14/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-