| Model Number |
TCN-10 |
| Medical Device Problem Codes |
Break (1069); Material Discolored (1170); Contamination /Decontamination Problem (2895)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/11/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The device analysis performed on the returned electrode revealed discolored epoxy and a chip out.The color of newly cured epoxy is amber, light brown.If the epoxy is subjected to too many autoclave cycles it becomes brittle and discolored.
|
| |
|
Search Alerts/Recalls
|