DEPUY SYNTHES SPINE ROD, 300 MM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Catalog Number 179762300
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Spinal Column Injury (2081)
Date of Event 01/01/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported on (b)(6), 2017, surgeon performed a case to treat a non union at l4/5.The patient had previously undergone a t10-ilium fusion.When speaking with surgeon, he informed me prior to the case that there was a non union at l4/5 which caused the rods to break.He exposed the fractured rods and used lateral connectors to place a second stabilizing rod on each side.In total, 6 lateral connectors were used (3 on each side).He also used and end to end connector to stabilize the rod at the fracture site.The case was completed successfully, and surgeon seemed pleased with the result.
 
Additional Manufacturer Narrative
(b)(4).Device was not returned for evaluation.A review of the device history record could not be performed as the lot number was not provided.A trend analysis was conducted.No emerging trends were found requiring further actions.Without the return of the device, we are unable to confirm the reported issue or identify the root cause.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.Should more information and/or the sample be provided at a later time, this complaint will be reopened and device evaluate.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROD, 300 MM
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
DEPUY SYNTHES SPINE
325 paramount drive
raynham MA 02767
MDR Report Key7118282
Report Number1526439-2017-50082
Device Sequence Number1119322
Product Code NKB
UDI-Device Identifier10705034077420
UDI-Public10705034077420
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K033901
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number179762300
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/29/2017
Supplement Date Received by Manufacturer01/03/2018
Initial Report FDA Received Date12/14/2017
Supplement Report FDA Received Date01/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-