| Model Number |
MICL13.7 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); No Code Available (3191)
|
| Date of Event |
09/08/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Work order search: no similar complaints types events were reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a micl13.7, -14.0 diopter, implantable collamer lens in the patient's right eye (od) on (b)(6) 2017.On (b)(6) 2017, the lens was explanted due to decrease visual acuity.At the time of submission, attempts to obtain additional information have been unsuccessful.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: product evaluation found that the lens was returned taped to an empty cartridge case.Visual inspection found debris on lens and the haptic missing a piece.(b)(4).The reporter indicated that the surgeon implanted a 13.7 mm micl13.7, -14.0 diopter, implantable collamer lens in the patient's right eye (od) on (b)(6) 2017.On (b)(6) 2017, the lens was explanted due to excessive vault, narrowing of the angle, and shallowing of the acd.The lens was exchanged for shorter lens and the problem was resolved.(b)(4).Conclusion code: - per dfu for this lens model, it is recommended only to use microstaar injectors, model msi-tf or msi-pf with sfc-45 or sfc-45 fp cartridge.(b)(4).
|
| |
|