|
Device evaluation summary: the visi-pro was received in post-inflation profile, the proximal end of the balloon chamber exhibited a radial tear, and the distal guidewire lumen exhibited stretching including tensional and compressional accordion folding.Not all components of the visi-pro delivery system were returned: the distal marker band, distal tip, and distal segment of the balloon chamber were not returned for evaluation.Approximately 137.3 cm of the working length of the delivery system was returned.Portions of catheter¿s working length exhibit tensile stretching.The proximal end of the balloon chamber was examined with the aid of a microscope.The proximal radiopaque marker band was present.The balloon chamber material exhibited radial tearing.The inner guidewire lumen exhibited tensile stretching and accordion folding.The distal radiopaque marker band, distal tip, and distal balloon chamber material were not returned for evaluation.If information is provided in the future, a supplemental report will be issued.
|