• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. RSP SHOULDER; SOCKET, SHELL NEUTRAL RSP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. RSP SHOULDER; SOCKET, SHELL NEUTRAL RSP Back to Search Results
Model Number 508-00-000
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032)
Date of Event 11/30/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient complaining of pain when moving shoulder.The surgeon took x-rays and there appeared to be an osteophyte inferior to the glenoid.Surgeon suspected it caused the prosthesis to impinge.During surgery, it was discovered that the prosthesis was able to be dislocated too easily.There was not enough tension so the surgeon had to build up thickness; discovered not impinging.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to pain.The previous surgery and the revision detailed in this investigation occurred over 5 years and 11 months apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was a significant adverse event to the patient.The healthcare professional indicated there was a 5 minute delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to the pain.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the event.There are multiple factors that may contribute to the event that are outside the control of djo surgical such as: aging of the patient, patient activities or trauma.The agent clearly stated that surgeon found osteophyte inferior to the glenoid through x-rays, hence patient bone deterioration may be the primary cause for pain.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP SHOULDER
Common Device Name
SOCKET, SHELL NEUTRAL RSP
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7120750
Report Number1644408-2017-01178
Device Sequence Number1799153
Product Code KWS
UDI-Device Identifier00888912024600
UDI-Public(01)00888912024600
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K041066
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/28/2017
Device Model Number508-00-000
Device Catalogue Number508-00-000
Device Lot Number848C1198
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/30/2017
Supplement Date Received by Manufacturer02/01/2018
Initial Report FDA Received Date12/15/2017
Supplement Report FDA Received Date02/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
508-00-032, LOT 855C1089
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age64 YR
-
-