• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number FUSION
Device Problem No Display/Image (1183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/17/2017
Event Type  malfunction  
Manufacturer Narrative
Patient information was unavailable from the site.A medtronic representative went to site to test the equipment.Representative was unable to replicate the reported issue.A system checkout was performed and the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.No parts were replaced.No parts have been received by the manufacturer for evaluation.The logs for the navigation system were reviewed by medtronic personnel.However, the logs provided no additional insight into the probable cause of the anomaly.
 
Event Description
A site representative reported that during functional endoscopic sinus surgery (fess), while navigation, the screen of the navigation system would not display any image.The procedure was completed with the use of navigation.There was a delay of less than 1 hour.No impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION NAVIGATION SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
prashanth gali
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7121736
MDR Text Key95074735
Report Number1723170-2017-05139
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994380005
UDI-Public00613994380005
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial
Report Date 12/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberFUSION
Device Catalogue Number9733560XOM
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/17/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/30/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-