• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problems No Display/Image (1183); Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 11/22/2017
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to perform a planned maintenance (pm). While on-site, the monitor of the navigation system was replaced. The hardware, software, and instruments passed the system checkout. The system was found to be fully functional. The monitor for the navigation system was returned to the manufacturer for evaluation. Testing found that the monitor would lose display functionality shortly after being plugged in. The hardware investigation found that the reported event was related to a hardware issue. This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Manufacturer Narrative
No patient information provided as no patient was involved in this concern. No parts have returned to the manufacturer for evaluation. No parts have returned for evaluation.
 
Event Description
A medtronic representative reported that while outside of a procedure, the surgeon monitor on the navigation system was black. It switches on briefly when connected, but then goes black. There was no patient present when this issue was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSTEALTHSTATION S7 SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7635267745
MDR Report Key7122050
MDR Text Key102487345
Report Number1723170-2017-05173
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeSE
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 03/02/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/15/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberS7
Device Catalogue Number9733858
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/13/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/10/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-