Model Number EGIA30AV
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/27/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter: occurred during a thoraco/lobectomy.During the second firing for right inferior lobe resection, the device could not be fired.The surgeon could not squeeze the handle.The case was completed using a new reload.No patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The reload was pre-fired and engaged in interlock.There were staples protruding from the proximal staple cartridge channel.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Replication of the pre-fire condition with interlock engagement may occur if the instrument firing handle had been partially compressed and released after pressing the green firing button.In this situation, the safety interlock feature will engage and prevent the reload from firing a second time by ceasing the placement of staples and tissue transection, and prevent patient harm.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN NORTH HAVEN - MFG
1995 mcdermott rd
north haven CT 06473
MDR Report Key7122880
Report Number1219930-2017-10005
Device Sequence Number13768155
Product Code GDW
UDI-Device Identifier20884521186092
UDI-Public20884521186092
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K111825
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2022
Device Model NumberEGIA30AV
Device Catalogue NumberEGIA30AV
Device Lot NumberN7E0655KX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/27/2017
Initial Date Received by Manufacturer 11/28/2017
Supplement Date Received by Manufacturer01/18/2018
Initial Report FDA Received Date12/15/2017
Supplement Report FDA Received Date01/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-