| Model Number |
RELTACK5R |
| Medical Device Problem Code |
Device Difficult to Setup or Prepare (1487)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/23/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic hernia procedure, the device was jammed on first firing.There were difficulties with loading the reload onto the handle, pressing the 'push to reload' button and navigating the lock and unlock button.Another device was opened with the same lot number and was used successfully.No patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: post market vigilance led an evaluation of one device.The loading unit was received fully fired with disrupted timing.Microscopic evaluation revealed scratches on the inner tube.Functionally, the dlu could not be loaded due to the disrupted timing of the device.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the reported conditions could be due to improper loading of the reloads, indicated by the scratches on the inner tube of the reloads.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|