CORDIS CORPORATION POWERFLEXPRO 6MM4CM 80; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
|
Back to Search Results |
|
| Model Number |
4400604S |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/19/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product was not returned for analysis.Device history record (dhr) review was conducted and the product met quality requirements for product acceptance.Additional information is pending and will be submitted within 30 days upon receipt.
|
| |
|
Event or Problem Description
|
|
During a leriche syndrome balloon kissing case, it was reported that the powerflex balloon catheter ruptured at its nominal pressure.There was no reported patient injury.It is unknown how the procedure completed, however, the procedure was successfully completed.The lesion was unknown.The patient¿s vessel level of tortuosity and calcification was unknown.The rate of stenosis was 100 % (cto).The product was clinically used and will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
During a leriche syndrome balloon kissing case, it was reported that the powerflex balloon catheter ruptured at its nominal pressure.There was no reported patient injury.It is unknown how the procedure completed, however, the procedure was successfully completed.The lesion was unknown.The patient¿s vessel level of tortuosity and calcification was unknown.The rate of stenosis was 100 %, chronic total occlusion (cto).The device was not returned for analysis.A device history record (dhr) review of lot 17689116 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst - at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause of the burst could not be determined.Based on the limited information available for review, vessel characteristics (100% stenosed lesion, cto) may have contributed to the reported event as calcification/resistant lesions may cause damage to a balloon.According to the instructions for use, which is not intended as a mitigation, ¿the rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over-pressurization.Do not exceed the rated burst pressure recommended on the label.Prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.¿ neither the dhr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective/preventive action will be taken at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received, the reportability determination remains unchanged.During a leriche syndrome balloon kissing case, it was reported that the powerflex balloon catheter ruptured at its nominal pressure.There was no reported patient injury.It is unknown how the procedure completed, however, the procedure was successfully completed.The lesion was unknown.The patient¿s vessel level of tortuosity and calcification was unknown.The rate of stenosis was 100 % (cto).There were no difficulty inflating the balloon during prep.The product was clinically used and will not be returned for analysis.No other information was provided.The device was not returned for analysis.A device history record (dhr) review of lot 17689116 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst - at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause of the burst could not be determined.Based on the limited information available for review, vessel characteristics (100% stenosed lesion, cto) may have contributed to the reported event as calcification/resistant lesions may cause damage to a balloon.According to the instructions for use, which is not intended as a mitigation, ¿the rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over-pressurization.Do not exceed the rated burst pressure recommended on the label.Prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.¿ neither the dhr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective/preventive action will be taken at this time.
|
| |
|