• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENSEAL 5MM ROUND TIP 35CM; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENSEAL 5MM ROUND TIP 35CM; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number NSEAL535RH
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/29/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch # unk.The lot history records were reviewed and the manufacturing criteria were met prior to the release of this lot.
 
Event or Problem Description
It was reported that during a total laparoscopic abdominal colectomy, after using on thick tissue, the jaws of the instrument wouldn't open.The doctor had to manually manipulate the jaws with his hands to open the jaws of the instrument.The device was removed from the patient and a second like device was used to complete the procedure.There were no patient consequences reported.
 
Additional Manufacturer Narrative
(b)(4).Batch # p9334a.Investigation summary: the device was received with no apparent damage.The device was tested on the generator and passed all functional testing.The energy output delivered from the device was verified.All three tones were heard during functional testing (tone 1 is heard when the energy activation button is pressed; tone 2 is heard when tissue impedance threshold is reached and tone 3 is heard when the cycle is complete).No issues were noted with opening and closing of the jaws.There were no anomalies noted with the functionality of the device.The batch history record was reviewed and no defects, nc¿s or protocols related to the complaint, were found during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSEAL 5MM ROUND TIP 35CM
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key7126328
Report Number3005075853-2017-06856
Device Sequence Number1358236
Product Code GEI
UDI-Device Identifier20705036000041
UDI-Public20705036000041
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K072177
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2019
Device Catalogue NumberNSEAL535RH
Device Lot NumberP9368F
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2018
Initial Date Received by Manufacturer 11/30/2017
Supplement Date Received by Manufacturer11/30/2017
Initial Report FDA Received Date12/18/2017
Supplement Report FDA Received Date01/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-