| Catalog Number |
NSEAL535RH |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/29/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # unk.The lot history records were reviewed and the manufacturing criteria were met prior to the release of this lot.
|
| |
|
Event or Problem Description
|
|
It was reported that during a total laparoscopic abdominal colectomy, after using on thick tissue, the jaws of the instrument wouldn't open.The doctor had to manually manipulate the jaws with his hands to open the jaws of the instrument.The device was removed from the patient and a second like device was used to complete the procedure.There were no patient consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # p9334a.Investigation summary: the device was received with no apparent damage.The device was tested on the generator and passed all functional testing.The energy output delivered from the device was verified.All three tones were heard during functional testing (tone 1 is heard when the energy activation button is pressed; tone 2 is heard when tissue impedance threshold is reached and tone 3 is heard when the cycle is complete).No issues were noted with opening and closing of the jaws.There were no anomalies noted with the functionality of the device.The batch history record was reviewed and no defects, nc¿s or protocols related to the complaint, were found during the manufacturing process.
|
| |
|