• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO PACIFIC XTREME; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO PACIFIC XTREME; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Catalog Number PCU070200130
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code Patient Problem/Medical Problem (2688)
Date of Event 11/27/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Physician employed the use of a pacific xtreme pta balloon catheter with a non-mdt inflation device for the treatment of a 100% stenosed fibrous/plaque lesion in the mid-proximal superficial femoral artery (sfa).Lesion was slightly calcified.Vessel slightly tortuous.Ifu was followed.Device prepped without issue.It was reported that a circumferential/radial balloon burst occurred at an inflation pressure of 8-9atms.One inflation was applied to the balloon previous to burst.The device did pass through a previously deployed stent, but no resistance was encountered / no excessive force required.A part of the balloon ruptured off the catheter and became lodged within the mid-sfa.A 10mm snare device was used to retrieve the ruptured balloon (all fragments retrieved).No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACIFIC XTREME
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
MEDTRONIC MEXICO
avenida paseo del cucapah #105
parque industrial el lago
tijuana,bc 22570
MX  22570
MDR Report Key7127035
Report Number9612164-2017-01998
Device Sequence Number1260569
Product Code LIT
UDI-Device Identifier08033477058320
UDI-Public08033477058320
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K103464
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2020
Device Catalogue NumberPCU070200130
Device Lot Number213333749
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/27/2017
Initial Report FDA Received Date12/18/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-