• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC OSCILLATING SAW ATTACHMENT II; BLADE, SAW, GENERAL AND PLASTIC SURGERY, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC OSCILLATING SAW ATTACHMENT II; BLADE, SAW, GENERAL AND PLASTIC SURGERY, SURGICAL Back to Search Results
Catalog Number 511.801
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/27/2017
Event Type  malfunction  
Manufacturer Narrative
Reporter number  (b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that during an unspecified surgical procedure, it was observed that the oscillating saw attachment device had an unspecified malfunction.It was unknown if there were any delays to the surgical procedure.It was reported that a spare device was available for use.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the device failed pretest for check the quick coupling for saw blades, and check oscillation frequency min 11,700 oscillation/minute.It was noted that the coupling tool side was not functioning.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to device use.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSCILLATING SAW ATTACHMENT II
Type of Device
BLADE, SAW, GENERAL AND PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES PRODUKTIONS GMBH
hauptstrasse 24
waldenburg 4437
SZ   4437
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 
6107195000
MDR Report Key7127761
MDR Text Key95615874
Report Number8030965-2017-50791
Device Sequence Number1
Product Code GFA
UDI-Device Identifier7611819102262
UDI-Public(01)7611819102262(11)120521
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 11/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/18/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number511.801
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/21/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/21/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-