MAUDE Adverse Event Report: WRIGHTS LANE : SYNTHES USA PRODUCTS LLC 11.0MM TI HELICAL BLADE 95MM-STERILE; ROD,FIXATION,INTRAMEDULLARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE : SYNTHES USA PRODUCTS LLC 11.0MM TI HELICAL BLADE 95MM-STERILE; ROD,FIXATION,INTRAMEDULLARY Back to Search Results
Model Number 456.304S
Medical Device Problem Code Sticking (1597)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/20/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device was used for treatment, not diagnosis.Patient weight and date of birth are not available for reporting.Additional product code hwc.Device was implanted on (b)(6) 2017 and is currently explanted.Explant date is non-applicable.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported a patient had initial surgery on (b)(6) 2017 for treatment of a sub-trochanteric femur fracture.Patient was implanted with one (1) trochanteric fixation nail (tfn), one (1) 11.0 mm ti helical blade and one (1) 4.5 mm distal locking screw.As the surgeon was using the slide hammer while inserting the helical blade implant with the helical blade coupling screw, the distal knob on the coupling screw broke off intraoperatively, and fell on the operating room floor.Due to the force that the surgeon used to disengage the helical blade implant from the tfn instruments, the aiming arm instrument and the helical blade inserter instrument became warped.Also, the buttress compression nut became stuck inside the blade guide sleeve.The surgeon successfully disengaged all the instruments from the implant, and proceeded with the surgery.No fragments were generated.Surgery was completed successfully with 30-40-minute time delay.Patient is reported in stable condition.This is report 4 of 4 for (b)(4).
 
Additional Manufacturer Narrative
Device is not expected to be returned for manufacturer review/investigation.Device was implanted on (b)(6) 2017 and is currently implanted.Explant date is non-applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
11.0MM TI HELICAL BLADE 95MM-STERILE
Common Device Name
ROD,FIXATION,INTRAMEDULLARY
Manufacturer (Section D)
WRIGHTS LANE : SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7127997
Report Number2939274-2017-50363
Device Sequence Number1256106
Product Code HSB
UDI-Device Identifier10886982226709
UDI-Public(01)10886982226709(17)EXPIRATIONUNKNOWN(10)LOTNUMBERUNKNOWN
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K011857
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 11/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number456.304S
Device Catalogue Number456.304S
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/20/2017
Supplement Date Received by Manufacturer01/17/2018
Initial Report FDA Received Date12/19/2017
Supplement Report FDA Received Date01/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age69 YR
-
-