| Model Number |
456.304S |
| Medical Device Problem Code |
Sticking (1597)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/20/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.Patient weight and date of birth are not available for reporting.Additional product code hwc.Device was implanted on (b)(6) 2017 and is currently explanted.Explant date is non-applicable.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported a patient had initial surgery on (b)(6) 2017 for treatment of a sub-trochanteric femur fracture.Patient was implanted with one (1) trochanteric fixation nail (tfn), one (1) 11.0 mm ti helical blade and one (1) 4.5 mm distal locking screw.As the surgeon was using the slide hammer while inserting the helical blade implant with the helical blade coupling screw, the distal knob on the coupling screw broke off intraoperatively, and fell on the operating room floor.Due to the force that the surgeon used to disengage the helical blade implant from the tfn instruments, the aiming arm instrument and the helical blade inserter instrument became warped.Also, the buttress compression nut became stuck inside the blade guide sleeve.The surgeon successfully disengaged all the instruments from the implant, and proceeded with the surgery.No fragments were generated.Surgery was completed successfully with 30-40-minute time delay.Patient is reported in stable condition.This is report 4 of 4 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device is not expected to be returned for manufacturer review/investigation.Device was implanted on (b)(6) 2017 and is currently implanted.Explant date is non-applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|