| Model Number |
60728753 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/09/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product analysis received and examined the returned product.Visual examination of the returned product confirmed the presence of a white, plastic-like, foreign material in the syringe barrel.The likely cause of the reported problem was determined to be the introduction of an unidentified particulate during component manufacturing.A 12 month review of complaint history determined the frequency to be (b)(4) complaints per million (cpm) for similar defects.
|
| |
|
Event or Problem Description
|
|
The customer reported the following: after opening the stellant ct disposable kit and upon inspection, they observed a white particulate inside the syringe barrel.The customer elected not to use the syringe kit.No injury or adverse event was reported.
|
| |
|