| Catalog Number |
CB11LT |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/29/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.
|
| |
|
Event or Problem Description
|
|
It was reported that during a laparoscopic nissen fundoplication, after inserting the trocar into the patient and inserting an instrument into the trocar, the doctor noticed a tiny piece of plastic dislodged as the instrument was inserted through the trocar, into the patient.The doctor was able to retrieve the tiny piece of plastic from the patient.The doctor continued to use the same trocar to complete the procedure.There were no patient consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # n92v1r, n92n8l.Device analysis: the analysis results found that the cb11lt device was returned with no damage in the external components.In addition, the obturator was not returned along with the instrument.The device was visually inspected and no missing pieces were observed from the sealing components.No conclusion could be reached as to what may have caused the reported incident.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
|
| |
|