DATASCOPE MAHWAH CS300; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Catalog Number 0998-00-3023-53
Device Problems Leak/Splash (1354); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(6).The production device history record (dhr) for this intra-aortic balloon pump (iabp) was not required to be reviewed per the company standard operating procedure since the device manufacture date is greater than one year from the event date.A company service territory manager (stm) evaluated the iabp and could not duplicate the alleged malfunction.The stm performed calibrations and function checkout.The iabp is functioning normally and was returned to the customer.
 
Event Description
The customer reported a cs300 intra-aortic balloon pump had a safety disk leak test failure.The circumstances under which the event occurred were not reported.No patient involvement or adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key7130909
MDR Text Key95805000
Report Number2249723-2017-01015
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 12/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0998-00-3023-53
Was Device Available for Evaluation? Yes
Date Manufacturer Received11/27/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-