| Catalog Number |
11-8999 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/20/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Returned for evaluation were the two locking screws replaced at the c2 level.A thorough visual examination was performed.Both units exhibited signs of wear consistent with surgical use.The locking screw threads displayed no abnormalities and functioned as intended upon evaluation.A review of the device history record found no non-conformances to specification.Based on the results of our evaluation, no definitive conclusion can be drawn.
|
| |
|
Event or Problem Description
|
|
At approximately 6 weeks post-operatively, it was noted on radiographs that two locking screws were dettached from the screw heads at the c2 level.A surgical procedure was performed at which time the locking screws at the c2 level were removed and replaced.The patient is recovering successfully.
|
| |
|