SIEMENS MEDICAL SOLUTIONS USA, INC. ACUSON Z6MS ULTRASOUND TRANSDUCER; ULTRASOUND DEVICE Back to Search Results
Model Number Z6MS TRANSDUCER
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 12/13/2017
Event Type  malfunction  
Manufacturer Narrative
The device referenced in this report has not been returned to siemens for evaluation.If additional information is received or if the device is returned at a later date, this report will be supplemented.(b)(4).
 
Event Description
It was reported that immediately upon initialization of the transducer, a usacquisitionhw_31 error message was displayed.The device was not being used on a patient at the time of failure.No additional information was provided.
 
Manufacturer Narrative
Investigation: the complaint was investigated for an acquisition 31 error when probe is initialized.The transducer was returned, and an investigation was performed.Engineers were able to reproduce an acquisition 31 error.The cause of the issue was determined to be a gastro flex thermistor trace fault due to gastro flex reliability; this is related to design.Improvements to the gastro flex trace were implemented into forward production, since july 2017.The transducer returned from the customer site was manufactured prior to the corrective action.The transducer was replaced at the site by service.Complaint reference #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACUSON Z6MS ULTRASOUND TRANSDUCER
Type of Device
ULTRASOUND DEVICE
Manufacturer (Section D)
SIEMENS MEDICAL SOLUTIONS USA, INC.
685 east middlefield road
mountain view CA 94043 4050
Manufacturer Contact
scott christiansen
685 east middlefield road
mountain view, CA 94043-4050
4252709862
MDR Report Key7131115
MDR Text Key95959754
Report Number3009498591-2017-00549
Device Sequence Number1
Product Code ITX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142628
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Biomedical Engineer
Remedial Action Modification/Adjustment
Type of Report Initial,Followup
Report Date 03/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/19/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberZ6MS TRANSDUCER
Device Catalogue Number10436113
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/13/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/14/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-