Catalog Number MSM20
Device Problem Failure to Form Staple (2579)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/28/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.
 
Event Description
It was reported that during a mastectomy the clips scissored on tissue on the first fire.The clips were easily removed from the tissue.The procedure was completed with another device with no patient consequences reported.
 
Manufacturer Narrative
(b)(4).Batch # p92u5w.The analysis results found that the msm20 device was returned with a clip in the jaws.  in an attempt to replicate the reported incident, the device was tested for functionality.Upon testing, the device was cycled and it fed and formed the remaining 4 clips as intended.As the device was found to be fully functional, it could not be determined what may have caused the reported incident.No conclusion could be reached as to what may have caused the reported incident.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key7133326
MDR Text Key95670971
Report Number3005075853-2017-06915
Device Sequence Number1
Product Code GDO
UDI-Device Identifier20705036002502
UDI-Public20705036002502
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K820837
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2022
Device Catalogue NumberMSM20
Device Lot NumberP4RP7F
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received12/21/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/26/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-