MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN@HOME, RF TELEMETRY BASIC PACEMAKER DATA TRANSMITTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN@HOME, RF TELEMETRY BASIC PACEMAKER DATA TRANSMITTER Back to Search Results
Model Number EX1150
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/16/2017
Event Type  malfunction  
Manufacturer Narrative
The results of the analysis confirmed the reported event of failure to start up. Inspection of the ac power adapter revealed burn marks on the main pcb consistent with damage due to the high current flow through the ac wall power outlet.
 
Event Description
This report is to advise of an event observed during analysis. Burn marks to the main pcb were revealed during analysis. There were no reported adverse patient consequences related to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMERLIN@HOME, RF TELEMETRY BASIC
Type of DevicePACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7133727
MDR Text Key95473196
Report Number2017865-2017-35687
Device Sequence Number1
Product Code DRG
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial
Report Date 12/20/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/20/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberEX1150
Other Device ID Number05414734504799
Was Device Available for Evaluation? No Answer Provided
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/04/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

-
-