MAQUET CARDIOPULMONARY AG OXYGENATOR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number BO-HQV 48412
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Information (3190)
Event Date 11/30/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A follow-up medwatch will be submitted when additional information becomes available.
 
Event Description
According to the hospital:"upon initiation of bypass, oxygenation of the blood was slow compared to what we normally see.The oxygenator did eventually oxygenate appropriately after ~1 minute." (b)(4).
 
Manufacturer Narrative
(b)(4).A follow-up medwatch will be submitted when additional information becomes available.Reference exemption # e2018002.(b)(4).The sample was investigated in the manufacturer laboratory.A quadrox-i neonatal was returned.The sample was contaminated.Purified with sodium hypochlorite.Leak test carried out according to lv 201.A leak could not be determined.Failure could not be confirmed.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.Since the reported failure did not contribute to a death or serious injury no corrective action is needed.In addition at this time it cannot be concluded that this is a systemic error.No corrective action is needed.
 
Event Description
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXYGENATOR, CARDIOPULMONARY BYPASS
Type of Device
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key7134070
MDR Text Key95801947
Report Number8010762-2017-00400
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K102464
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/06/2018
Device Model NumberBO-HQV 48412
Device Catalogue Number701054973
Device Lot Number92213655
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/30/2017
Initial Date FDA Received12/20/2017
Supplement Dates Manufacturer Received11/30/2017
Supplement Dates FDA Received05/17/2018
Date Device Manufactured10/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-