BOSTON SCIENTIFIC - COSTA RICA (COYOL) AUTOTOME¿ RX 39; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number M00545200
Device Problem Failure to Align (2522)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/21/2017
Event Type  malfunction  
Manufacturer Narrative
The device has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that an autotome¿ rx 39 was used during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2017.According to the complainant, during the procedure, it was difficult to bow the autotome.When the physician was finally able to bow the autotome, the device stayed in the bowed position.The procedure was completed with another autotome¿ rx 39.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Manufacturer Narrative
A visual evaluation of the returned device found that the working length was twisted and the cutting wire was kinked.The exposed cutting wire length was within specification.Functional inspection was performed.The unit was tested inside and outside the duodenoscope, and the unit was able to bow and release the bow within specification.The condition of the returned unit was not consistent with the complaint incident that the tip of the tome failed to unbow.The complaint was not confirmed.Therefore, the most probable cause is operational context, since anatomical and/or procedural factors encountered during procedure could have affected the device performance.A dhr (device history record) review was performed and no deviation was found.A search of the complaint database confirmed that no other complaints exist for the specified batch.
 
Event Description
It was reported to boston scientific corporation that an autotome rx 39 was used during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2017.According to the complainant, during the procedure, it was difficult to bowthe autotome.When the physician was finally able to bow the autotome, the device stayed in the bowed position.The procedure was completed with another autotome rx 39.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOTOME¿ RX 39
Type of Device
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS  
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key7134831
MDR Text Key95635777
Report Number3005099803-2017-03673
Device Sequence Number1
Product Code KNS
UDI-Device Identifier08714729444770
UDI-Public08714729444770
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K013153
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/07/2020
Device Model NumberM00545200
Device Catalogue Number4520
Device Lot Number21110309
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/22/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/08/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-