Catalog Number 0998-00-3023-53
Device Problems Failure to Power Up (1476); Battery Problem (2885)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/28/2017
Event Type  malfunction  
Manufacturer Narrative
A getinge service territory manager (stm) confirmed the iabp had no ac power.The stm checked all fuses, the power cord, and switches.All of them were ok.The stm replaced the power supply and the problem resolved.The iabp passed at safety testing and was cleared for clinical use.The production device history record (dhr) for this intra-aortic balloon pump (iabp) was reviewed and no non-conformances related to the reported event were noted.
 
Event Description
It was reported that the cs300 intra-aortic balloon pump (iabp) has no ac power, and the iabp is always running on battery.No patient involvement or adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key7135196
MDR Text Key95734810
Report Number2249723-2017-01038
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 12/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 11/29/2017
Initial Date FDA Received12/20/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/02/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-