• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ACE SURGICAL SUPPLY COMPANY INTERNAL HEX IMPLANT; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ACE SURGICAL SUPPLY COMPANY INTERNAL HEX IMPLANT; DENTAL IMPLANT Back to Search Results
Model Number 205111
Medical Device Problem Codes Failure to Osseointegrate (1863); Osseointegration Problem (3003)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 09/03/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported event has been determined to be a post-placement/pre-load failure.This implant loss suggests that the source of the problem was likely to stem from procedural errors and/or the lack of osseointegration.Additionally, other factors that may have contributed to the implant failure includes bone condition, patient oral hygiene, or patient behavior.Proper intended use of the implant is described in its instructions for use.
 
Event or Problem Description
Mobility and inadequate bone quality/bone quantity were reported.Bone quality at the time of implant failure type iii.Time of loss in relation to the implantation was before prosthetic restoration.Primary stability was achieved.Osseointegration was not achieved.Augmentation procedure was not performed.Patient is a smoker.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERNAL HEX IMPLANT
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ACE SURGICAL SUPPLY COMPANY
1034 pearl street
brockton MA 02301
MDR Report Key7135272
Report Number1287163-2017-51258
Device Sequence Number864245
Product Code DZE
UDI-Device Identifier00614950004812
UDI-Public00614950004812
Combination Product (Y/N)N
Initial Reporter CountryLE
PMA/510(K) Number
K102981
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 11/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Dentist
Device Model Number205111
Device Catalogue Number205111
Device Lot Number13020046
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2017
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/30/2017
Initial Report FDA Received Date12/20/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-