Model Number H749A70200
Device Problem Retraction Problem (1536)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
It is indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.(b)(4).
 
Event Description
Same case as mdr id: 2134265-2017-12972 and 2134265-2017-12970.It was reported that automatic pullback failure occurred.A 100v ilab ultrasound imaging system was used in conjunction with an unknown imaging catheter and pullback sled to perform automatic pullback.During the procedure, it was reported that the ilab ultrasound imaging system froze frequently during pullback.The issue was resolved by correcting the contact or connection of the internal board of the acquisition processor of the ilab ultrasound imaging system.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALAXY INTRAVASCULAR ULTRASOUND SYSTEM
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BOSTON SCIENTIFIC - FREMONT (SUD)
47215 lakeview blvd phone
west dock
fremont CA 94538
Manufacturer (Section G)
BOSTON SCIENTIFIC - FREMONT (SUD)
47215 lakeview blvd phone
west dock
fremont CA 94538
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7135343
MDR Text Key95667982
Report Number2134265-2017-12971
Device Sequence Number1
Product Code IYO
Combination Product (y/n)N
PMA/PMN Number
K980851
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 12/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberH749A70200
Device Catalogue NumberA7020
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/08/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N