• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS; TORIC IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS; TORIC IOLS Back to Search Results
Model Number ZCT525
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Visual Impairment (2138); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2017.(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a zct525 24.0 diopter intraocular lens (iol) was implanted into the patient''s operative eye on (b)(6) 2017.It was later explanted on (b)(6) 2017, because the patient wasn''t seeing well, probably due to an inappropriately chosen lens.There was no incision enlargement, vitrectomy, or sutures used.The replacement lens was a zct225 23.5 diopter iol.Reportedly, there was no patient injury and the status of the patient post-operatively was unknown.No additional information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation: yes.Returned to manufacturer on: 01/22/2018.Device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection with the unaided eye shows the product is cut in half, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA 92705
MDR Report Key7136138
Report Number9614546-2017-01249
Device Sequence Number13815607
Product Code HQL
UDI-Device Identifier05050474603653
UDI-Public(01)05050474603653(17)201022
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physical Therapist
Device Expiration Date10/22/2020
Device Model NumberZCT525
Device Catalogue NumberZCT525U240
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2018
Initial Date Received by Manufacturer 11/21/2017
Supplement Date Received by Manufacturer02/09/2018
Initial Report FDA Received Date12/20/2017
Supplement Report FDA Received Date03/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-