• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SENSOR ENLITE MMT-7008A; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SENSOR ENLITE MMT-7008A; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-7008A
Device Problems Device Displays Incorrect Message (2591); Device Operates Differently Than Expected (2913)
Patient Problems Bruise/Contusion (1754); Blood Loss (2597)
Event Date 05/06/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call the sensor had inaccurate readings that triggered threshold suspend alarm.The customer¿s blood glucose was 220 mg/dl and the sensor glucose was 60 mg/dl at the time of the incident.Customer stated sensor was curved.Sensor glucose value that triggered the suspend event was 60 mg/dl.Suspend on low limit in sensor settings was 80 mg/dl.The customer stated that customer sensor reading were low and threshold was suspending.The sensor will be returned for analysis.
 
Manufacturer Narrative
Inspected one opened and used sensor and performed continuity resistance test and sensor passed per specifications.Performed bicarbonate buffer test and sensor passed per specifications.Found cannula bent unable to confirm customer received sensor in said condition due to customer return sensor opened and used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSOR ENLITE MMT-7008A
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7138654
MDR Text Key95509980
Report Number2032227-2017-75697
Device Sequence Number1
Product Code OZO
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 06/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-7008A
Device Catalogue NumberMMT-7008A
Device Lot NumberHG1QCE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2018
Date Manufacturer Received05/29/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age32 YR
Patient Weight132
-
-