| Model Number |
0180 |
| Medical Device Problem Code |
Installation-Related Problem (2965)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
12/04/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Incorrect installation.The installer did not follow the instructions and lifted the table without transportation clamp mounted.The table top fell down.No person was injured.Since this is the second time (happened once in 2015 for an oem customer) we consider this as a foreseeable misuse.As an immediate action the tables will be delivered with a warning sign inside the crate.We will change the design in order for the table to withstand this type of incorrect handling.Design change will be fully implemented in february 2018.Already installed tables are safe to use.No field action is required.
|
| |
|
Event or Problem Description
|
|
Happened during installation.The installer did not follow the instruction and lifted the table without the safety clamp mounted.The table top fell down.No person was injured.
|
| |
|