• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. KIT MMT-1760G 670G REVENUE BLACK MG; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. KIT MMT-1760G 670G REVENUE BLACK MG; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL Back to Search Results
Model Number MMT-1760K
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/08/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event or Problem Description
The customer reported via phone call that the the insulin pump continue to alarm and had the symbol of the red arrow pointing to a manual.The customer's blood glucose was 135 mg/dl.The customer stated that the fluid under pump.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will not be returned for analysis.
 
Additional Manufacturer Narrative
Insulin pump was received with critical pump error (open book image) alarm and unable to download due to moisture damage on the electronic assembly.Unable to perform the self-test, displacement test, rewind test, prime/seating test, basic occlusion test, force sensor test, occlusion test, sleep current measurement test, and active current measurement test due to critical pump error (open book image) alarm.Moisture damage was also found on the force sensor and motor during visual inspection.No moisture damage was found on the keypad assembly.Insulin pump was received with partially broken reservoir tube lip, missing retainer, missing reservoir tube o-ring, scratched case, case cracked behind the insulin pump at the corner of the belt clip rails, missing display window cover, pillowing keypad overlay, and minor scratched lcd window.
 
Additional Manufacturer Narrative
This report is part of a retrospective review and remediation efforts in response to a warning letter.Updated h9: z-0956-2020.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KIT MMT-1760G 670G REVENUE BLACK MG
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7139226
Report Number2032227-2017-75764
Device Sequence Number10997827
Product Code OZP
UDI-Device Identifier00643169946040
UDI-Public(01)00643169946040
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Patient
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-1760K
Device Catalogue NumberMMT-1760K
Device Lot NumberHG1R0MQ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2018
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 12/08/2017
Supplement Date Received by Manufacturer02/07/2018
10/14/2019
Initial Report FDA Received Date12/21/2017
Supplement Report FDA Received Date02/21/2018
06/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0956-2020
Patient Sequence Number1
Patient SexUnknown
-
-