It was reported during an initial implant surgery that upon interrogation, the new generator was found to be depleted.The surgeon mentioned that prior to implant, he did not recall the device being depleted and that the end of service indicator was observed after multiple interrogations and diagnostics.Impedance of the new generator was within normal limits.A backup generator was used for the implant.A review of device history records revealed that the generator passed quality control inspection prior to distribution.The explanted device has not been received by the manufacturer to-date.No additional relevant information was received to date.
|
Product analysis was performed and approved for the explanted generator.Burn marks were observed on the generator case, indicating that the generator may have been exposed to an electrocautery tool during the implant procedure.The minimum voltage measured showed 1.786 v, estimated from the day of surgery, confirming the report of the "end of service" indicator.Decoded data from the generator indicated that 1.090% of the battery had been consumed.Battery voltage measured 2.964 during testing, indicating an intensified follow-up indicator = no (ifi=no) indicator.Testing performed on the device showed the generator performed according to functional specifications.Other than the noted "end of service" event, there were no additional performance or any other type of adverse conditions found with the generator.
|